West Virginia Bids > Bid Detail

J043--Medical Air Systems Maintenance-581

Agency: VETERANS AFFAIRS, DEPARTMENT OF
Level of Government: Federal
Category:
  • J - Maintenance, Repair, and Rebuilding of Equipment
Opps ID: NBD00159761104647667
Posted Date: Nov 14, 2022
Due Date: Nov 23, 2022
Solicitation No: 36C24523Q0091
Source: https://sam.gov/opp/c0d325fce1...
Follow
J043--Medical Air Systems Maintenance-581
Active
Contract Opportunity
Notice ID
36C24523Q0091
Related Notice
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
245-NETWORK CONTRACT OFFICE 5 (36C245)
Looking for contract opportunity help?

Procurement Technical Assistance Centers (PTACs) are an official government contracting resource for small businesses. Find your local PTAC (opens in new window) for free government expertise related to contract opportunities.

General Information
  • Contract Opportunity Type: Combined Synopsis/Solicitation (Updated)
  • All Dates/Times are: (UTC-05:00) EASTERN STANDARD TIME, NEW YORK, USA
  • Updated Published Date: Nov 14, 2022 02:12 pm EST
  • Original Published Date: Nov 02, 2022 12:26 pm EDT
  • Updated Date Offers Due: Nov 23, 2022 04:00 pm EST
  • Original Date Offers Due: Nov 16, 2022 04:00 pm EST
  • Inactive Policy: Manual
  • Updated Inactive Date: Dec 23, 2022
  • Original Inactive Date: Dec 16, 2022
  • Initiative:
    • None
Classification
  • Original Set Aside:
  • Updated Set Aside:
  • Product Service Code: J043 - MAINT/REPAIR/REBUILD OF EQUIPMENT- PUMPS AND COMPRESSORS
  • NAICS Code:
    • 811210 - Electronic and Precision Equipment Repair and Maintenance
  • Place of Performance:
    Huntington VAMC 1540 Spring Valley Road Huntington , WV 25704
Description
Solicitation Question; Answered by Jim Adkins, M&O Foreman, ext. 3383

1. Are we able to schedule a site visit for this? Yes, please contact Jim Adkins

2. Preventative Maintenance on all hospital bed head units where oxygen, med gas, and vacuum/suction connections are located as needed. There isn't a PM schedule for these that I am aware of. We can test outlets and check for leaks. Does this suffice?

That is fine, however; contract to include repairs/parts as needed

3. Vendor to provide all parts, labor, material, equipment, and supervision too conduct Quarterly Preventive Maintenance on three Medical Air Systems, three Medical Vacuum Systems and all hospital bed head units where oxygen, med gas and vacuum/suction connections are located. PM pricing on pumps and compressors can be provide but ca the others be quoted as needed.

Sure

4. The PM compressor equipment info seems to not be the system manufacturers info. Can you please provide correct information?

Already provided; please contact Jim Adkins for further info.

5. There is a 2 hour response time in the SOW, can this be amended to reflect a 48-hour response time?

No
Attachments/Links
Contact Information
Contracting Office Address
  • 8849 International Drive Suite 215
  • LINTHICUM , MD 21090
  • USA
Primary Point of Contact
Secondary Point of Contact


History

TRY FOR FREE

Not a USAOPPS Member Yet?

Get unlimited access to thousands of active local, state and federal government bids and awards in All 50 States.

Start Free Trial Today >